CLIA Kit for Selectin, Endothelium (SELE)

Cloud-Clone Corp.
Product Code: SCA029Hu
Product Group: CLIA Kits
CodeSizePrice
SCA029Hu-24T24T£254.00
Quantity:
SCA029Hu-48T48T£422.00
Quantity:
SCA029Hu-96T96T£589.00
Quantity:
SCA029Hu-96T*596T*5£2,544.00
Quantity:
SCA029Hu-96T*1096T*10£4,779.00
Quantity:
Prices exclude any Taxes / VAT

Overview

Regulatory Status: RUO
Target Species: Human
Application: Chemiluminescent Immunoassay (CLIA)

Images

1 / 1

Further Information

Alternative Names:
CD62E; CD62-E; E-Selectin; ELAM; ELAM1; ESEL; SEL-E; E-LAM; E-LAM1; E-SEL; LECAM2; Endothelial Leukocyte Adhesion Molecule 1; CD62 Antigen-Like Family Member E
Assay length:
2h, 40min
Assay Procedure Summary:
1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 2 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
8. Read RLU value immediately.
Detection range:
6.86-5,000pg/mL
Format:
48T, 96T, 96T?5, 96T?10, 96T?100
Item Name:
Selectin, Endothelium
Method:
Double-antibody Sandwich
Precision:
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level E-selectin were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level E-selectin were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
Research Area:
Signal transduction;CD & Adhesion molecule;Tumor immunity;Infection immunity;
Sample type:
Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Sensitivity:
The minimum detectable dose of this kit is typically less than 2.63pg/mL
Specificity:
This assay has high sensitivity and excellent specificity for detection of E-selectin.
No significant cross-reactivity or interference between E-selectin and analogues was observed.
Stability:
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
Test Principle:
The microplate provided in this kit has been pre-coated with an antibody specific to E-selectin. Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to E-selectin. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is proportional to the E-selectin level in the sample or standard.;
Uniport ID:
P16581